New TopShim run:

version = topshim 2.0.0
date = 2025-06-07
time = 11:03:39.400
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0631
solvent = CD3CN
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 1.93 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

initial B0 stdDev = 0.25 Hz
sample size = 3.00 cm, position = -0.02 cm
final B0 stdDev = 0.23 Hz > improvement = 1.1
envelope width = 0.45 Hz

Performing tune after shimming
tune lock level increase = 1.1

shim changes: 
Z	+10
Z2	+1
Z3	-19
Z4	-17
Z5	+39
X	-10
Y	-11

duration = 2 min 24 sec
completed successfully
finished Sat Jun  7 11:06:00 2025


========================================================================